Diagnostic Imaging Physics
MPHY-501
Bushberg Chapter 6
X-Ray Production, X-Ray Tubes and X-Ray Generators
During the next two years you will learn the basics of Diagnostic Imaging. Ultimately after this period you will enroll in a Medical Physics Residency Program and finally take the Board examination that, once you pass, will certify you as a medical physicist.
1
Basic Components
X-Ray Tube Insert
Electron Source
Medium for Acceleration (vacuum)
Electron Rich Target
X-ray Generator
Voltage Source to Create Potential Difference
Current
Exposure Timer
Tube Housing
Protective Shielding
Cooling
Tube Port
Filters (Beam Quality)
Collimators
X-Ray Production Process
Stage 1: Heat
Heat is required to boil electrons off of the Cathode
Stage 2: Acceleration
A potential difference (voltage) is placed across the Cathode and Anode
Stage 3: X-Ray production
Accelerated electron collides with the Anode (the Target)
Various atomic and nuclear interactions occur causing:
The initial accelerated electron to decelerate and give off x-rays
Electrons from target to be:
Ejected from targets atoms giving off x-rays
Change energy levels (orbits) on the targets atoms, giving off x-rays
Nuclear Collision
Electron capture (or similar event) causing total loss of electrons kinetic energy, giving off a g-Ray
e-
e-
e-
e-
e-
e-
e-
e-
e-
e-
e-
e-
e-
e-
e-
e-
e-
ANODE
+
Focusing Cup
Bias Voltage
Heater
+ High Voltage –
e-
e-
e-
e-
e-
e-
e-
e-
e-
e-
e-
e-
e-
X-Ray
X-Ray
X-Ray
120,000 Volts
CATHODE
*Focusing Cup is Shaped like the letter C and Cathode starts with the Letter C
Concept
Kinetic Energy of Electrons are Transferred to X-Rays
Several types of electron interaction x-ray spectrum is continuous
0
20
40
60
30
50
70
80
90
10
Relative Output
Energy (keV)
Average Energy is 1/3 to 1/2 Maximum Energy
Filtered Bremsstrahlung Spectrum
Unfiltered Bremsstrahlung Spectrum
Filtration:
Objects like glass x-ray tube
Other objects purposely placed in beam like aluminum
Ratio of Radiative (Bremsstrahlung) energy loss to collisional energy loss
From 0 to 150keV
Several Levels of Energy Interactions 0 – >.5MeV
Bremsstrahlung radiation happens when electro changes direction this in order to balance the energy equation
Conduction band and Fluorescence band electrons
Inner shell electron interactions
Nuclear Attraction
Nuclear Absorption
Energetic Interactions
+
n
+
n
+
n
E0
l2
l1
Esc
q
Compton Electron
(Ee-)
Compton
e-
e-
e- + p = n + g
Non Schell Electrons
Bremsstrahlung
Increasing Energy
Decreasing Energy
e-
7
Characteristic X-ray generation
+
n
+
n
+
n
X-Ray emitted
Due to electrons
energy shell change
Ejected K-Shell elecrton
Incoming Electron
Rebound Electron
Adjacent shell electron filling in
8
W, 90keV, Characteristic Spectrum
0
20
40
60
30
50
70
80
90
10
Relative Output
Energy (keV)
Filtered Bremsstrahlung Spectrum
Ka1
Ka2
Kb1
Extra Credit:
Name this peak
Shells and Subshells Filling
Shell name Subshell name Subshell max electrons Shell max electrons
K 1s 2 2
L 2s 2 2 + 6 = 8
2p 2p1/2 2p3/2 6
M 3s 2 2 + 6 + 10 = 18
3p 3p1/2 3p3/2 6
3d 3d1/2 3d3/2 10
N 4s 2 2 + 6 +10 + 14 = 32
4p 4p1/2 4p3/2 6
74 W
69.525
12.100
11.544
10.207
2.820
2.575
2.281
1.872
1.809
594.1
490.4
423.6
Element
K 1s
L1 2s
L2 2p1/2
L3 2p3/2
M1 3s
M2 3p1/2
M3 3p3/2
M4 3d3/2
M5 3d5/2
N1 4s
N2 4p1/2
N3 4p3/2
10
Binding Energies and Characteristic Shell Edges
Electron Binding Energies keV of Common X-Ray Tube Targets
Electron Shell Tungsten Molybdenum Rhodium
K 69.5 20.0 23.2
L 12.1 11.5 10.2 2.8 2.6 2.5 3.4 3.1 3.0
M 2.8 1.9 0.5 0.4 0.6 0.2
K-Shell Characteristic X-Ray Energies (keV) of Common X-Ray Tube Targets
Shell Transition Tungsten Molybdenum Rhodium
Ka1 59.32 17.48 20.22
Ka2 57.98 17.37 20.07
Kb1 67.24 19.61 22.72
Basic Components
X-Ray Tube Insert
Electron Source (Cathode)
Medium for Acceleration (vacuum)
Electron Rich Target (Anode, can be a rotor/stator or fixed)s
X-ray Generator
Voltage Source
Potential Difference,
Mammography 25-40keV,
Other imaging 40-150keV
Current
Fluoroscopy 1-5 mA in continuous mode, 10-50mA in pulsed mode
Projection Radiography 50-1,2000 mA,
1mA = 6.25 X1015 electrons/sec
Exposure Timer
In projection Radiography typically < 10ms Tube Housing Protective Shielding Cooling Tube Port Filters (Beam Quality) Collimators Many times the product of Tube current and exposure time are combined into one unit: milliampere Seconds (mAs) Cathode The Cathode is the negatively charged electrode in the x-ray tube Combination of: filament focusing cup Filament receives ~10-Volts and up 7 ampere Most x-ray tubes have two filaments, one small, one large Each filament is placed into its own machined focusing cup One end of filament may be connected to or insulated from the focusing cup The other end of the filament is definitely insulated from the focusing cup Only one of the two filaments is energized at any given time Current from the filament circuit heats a filament, which releases electrons by thermionic emission Focusing Cups At the top are typical electron distributions incident on the target anode (the focal spot) for the unbiased and biased focusing cups. Application of ?4,000 V on an isolated focusing cup completely stops electron flow, even with high voltage applied on the tube; this is known as a grid biased or grid pulsed tube The focusing cup shapes the electron distribution when it is at the same voltage as the filament Isolation of the focusing cup from the filament and application of a negative bias voltage (? .100 V) reduces electron distribution further by increasing the repelling electric fields surrounding the filament and modifying the electron trajectories Tube Current Vs. Filament Current Dependence of approximately kV1.5 For tube voltages 40 kV and lower, a space charge cloud shields the electric field so that further increases in filament current do not increase the tube current. This is known as space charge-limited operation Above 40 kV, the filament current limits the tube current; this is known as emission-limited operation. Anodes Heat destroys anodes Damaged Anode produce fewer x-rays This requires more current that then causes more heat https://www.spellmanhv.com/en/Technical-Resources/Application-Notes-X-Ray-Generators/AN-02 Anodes Tungsten 90%/Rhenium10% alloy is good choice due to heat profile W Melting Point = 3,422C = 3,695K = 6,192F = (darn hot) To prevent heat damage the duration of x-ray production is limited Other Anodes are used to take advantage of characteristic x-rays Molybdenum Rhodium See Chapter 8 Essentials of Medical Physics (Third Edition), Bushberg, et.al. Anode Configurations Fixed Dental Mobile devices Some Portable Fluoroscopy Rotating Comprises most devices for heat management reasons Can withstand higher intensity production Rotation speeds 3,000 3,600 RPM low speed 9,000 10,000 RPM High Speed System is designed to not allow electron flow until anode is at full speed This is why these systems require one or more seconds before exposure button is pushed For an excellent description of rotating-anode failure modes see: https://www.spellmanhv.com/en/Technical-Resources/Application-Notes-X-Ray-Generators/AN-02 Simple Fixed Anode The anode of a fixed anode x-ray tube consists of a tungsten insert mounted in a copper block. Heat is removed from the tungsten target by conduction into the copper block. Rotating Anode The anode of a rotating anode x-ray tube is a tungsten disk mounted on a bearing-supported rotor assembly. The rotor consists of a copper and iron laminated core and forms part of an induction motor. Stator, which exists outside of the insert. A molybdenum (poor heat conductor metal) stem connects the rotor to the anode to reduce heat transfer to the rotor bearings (bottom) Focal Spot The anode (target) angle, ?, is angle of the target surface wrt the central ray. The focal spot length, as projected down the central axis, is foreshortened, according to the line focus principle (lower right). Effective Focal length = Actual Focal Length X sin q Diagnostic Imaging: 7o < q < 20o Commonly 12o < q < 15o Focal Length Examples EXAMPLE 1: Actual anode focal area is 4 mm (length) by 1.2 mm (width). Anode angle is 20-degrees, What is projected focal spot size at central axis position? Answer: Effective length = actual length × sin ? = 4 mm × sin 20 degrees = 4 mm × 0.34 = 1.36 mm; therefore, the projected focal spot size is 1.36 mm (length) by 1.2 mm (width). EXAMPLE 2: If Example 1s anode angle is reduced to 10 degrees and actual focal spot size remains same, what is projected focal spot size at central axis position? Answer: Effective length = 4 mm × sin 10 degrees = 4 mm × 0.174 = 0.69 mm; thus, the smaller anode angle results in a projected size of 0.69 mm (length) by 1.2 mm (width) for the same actual target area. CATHODE e- e- e- e- e- e- e- e- e- e- e- e- Heal Effect Effective Focal Spot: Terminology Anode Side Cathode Side Heal Effect Anode Anode Side of the Anode Cathode Side of the Anode Cathode Side of Imaging Plane Anode Side of Imaging Plane More Photons Fewer Photons Heal Effect is Less Prominent at large Source Image Distance (SID) Source distance is measured from focal spot center of anode Image distance measured from center of imaging plane SID Differences in Projected Focal Spot Size Within the Imaging Field Anode Side Cathode Side NEMA Standards for Focal Spot Sizes There are three limits, small, medium and large focal spots Focal spots can be larger than nominal size but NOT SMALLER. For focal spots Must be done at 74keV and 50% of max mA for each spot size (remember that many cathodes have two elements) fs < 0.8 mm + 50% 1.5mm ? fs ? 0.8mm +40% fs > 1.5mm +30%
Start your Spread Sheets!!!!!!
Focal Spot Size Focal Spot Size
Nominal Size Tested Size PASS/FAIL Passing Criteria
fs < 0.8 mm + 50%
1.5mm ? fs ? 0.8mm +40%
fs > 1.5mm +30%
4 Ways to test focal spot size
Pin hole
Typically 10micrometer to 30micrometer diameter hole (highly attenuating)
Slit
Must be placed in two different directions
Along the anode cathode direction
Perpendicular to the anode cathode direction
Star
Bar
How many lines/mm is a resolution of 0.6mm?
A line pair consists of a dark line and a bright line.
So if one line is 5 microns wide, then a line pair will be 10 microns wide and there would be 1 mm/10 microns = 100 line pairs per millimeter.
Off-Focal Radiation
Electrons scatter off or anode are re-accelerated to anode but not to the focal spot
Give low intensity x-rays about the face of the anode
Many off-focal radiation are caught by the collimator
Grounded metal X-ray tube enclosures attract many of the scattered electrons thus keeping them from returning to anode
Off-Focal Radiation Distribution
Hot Spots
Description: X-Ray Tube Insert
Contains
Cathode
Anode
rotor assembly
sealed support structures
glass or
metal
enclosure under a high vacuum
Gas molecules leach from internal structures
Getter eliminates those molecules
Note1: Getter is a deposit of reactive material placed inside a vacuum system, for completing and maintaining the vacuum. When gas molecules strike the getter material, they combine with it chemically or by absorption. Thus the getter removes small amounts of gas from the evacuated space.
https://en.wikipedia.org/wiki/Getter#:~:text=A%20getter%20is%20a%20deposit,gas%20from%20the%20evacuated%20space.
Note 2:
Mammography Inserts include a beryllium (z=4) window in the port specifically to minimize the the absorption of low energy x-rays.
Other ports (non mammography) use the same material as the tube itself.
Description: X-Ray Tube Housing
Between Insert an Housing is Oil, provides:
Electrical insulation
Heat conduction (cooling)
Expansion bellows (for the expansion of oil as it heats up)
Overtemperature (senses oil temperature) switch stops x-ray production
CT and Interventional Fluoroscopy (high duty cycle) Housings
may have built-in heat exchangers (for cooling)
Shielding for scattered leakage (x-rays not in direction of imaging)
21 CFR (code of federal regulations) 1020.30 requires
leakage less than 100 mR/hr (0.88 mGy/hr) @ one meter from focal spot
Leakage must be tested at maximum rated kV (typically 125kV – 150kV) and highest possible current typically 3mA to 5mA
Equipment software/hardware should not allow operation at above max rated kV
Testing with
Area survey dose rate meter
Collimators
Adjust size and shape of x-ray field from tube port
Parallel opposed shutters
rectangular FOV for plain film
other shapes available (Fluoroscopy, hexagonal Iris )
Collimator Assembly
Lead insert Shield
Collimator
Assembly
Collimation must be within 2% of SID
Example: SID = 100cm
Collimation must be:
100cm ± 2cm, 40 ± 0.8
20CFR1020.31
Positive Beam Limitation (PBL)
Automatically senses film cassette size and adjusts so that beam cannot exceed cassette size
Changeable Filter
Al, Cu, Moly, Rho,
Special Tube Designs
Pulsed Fluoroscopy
Pro: 4kV grid bias keeps electrons in cloud by cathode when off electrons rapidly fly toward anode allowing rapid pulse times.
Con: Costs much more $$$$$
Mammography
Special tube output port (Beryllium instead of glass or other metal)
Potentially Different Targets (molybdenum, rhodium, tungsten)
Smaller Focal Spot (0.3mm, 0.1mm)
Grounded Anodes
CT
High instantaneous output, high heat load, Rapid Cooling
Rapid Tube Rotation about Gantry. More than 200RPM!!! https://www.youtube.com/watch?v=aBlJebipLgM
CT Advanced X-Ray Tube Features:
Anode & planar cathode within rotating vacuum enclosure.
Bearings mounted outside vacuum enclosure.
Deflection coils magnetically direct electron beam to specific areas on the target.
Circulating oil rapidly removes excess heat from the anode.
Flying Focal Spot: Electron beam rapidly deflects between two focal spots
How to Maximize X-Ray tube life
Hold ready state only as long as needed (dead man 2-stage switch)
Holding on keeps current in filament damages filament
Evaporated tungsten deposits on inside of x-ray tube (not good)
Use lower tube current and longer exposure times (when reasonable)
Avoid extended repeat operation at high technique
Can etch anode and cause several problems lower photon output etc.
Do not make high mA exposure on cold anode
follow manufactures recommendations
Limit Rotor start & Stop operations
Allow 30 to 40 seconds between exposures
Filtration
Inherent Filtration
X-ray Tube
Glass (SiO2) and Aluminum have similar attenuations, ZSiO2=14, Zal = 13
Glass and Aluminum attenuate all x-rays below 15 keV
Port
Housing Oil
Collimator Assembly
Mirror
Added (beam quality filters)
Intended to eliminate low energy x-rays that add to dose but not image
Metal
Plastic
X-Ray Generators Voltage Needs
Np = 5
Ns = 10
Vs = 2Vp
Transformer Nomenclature| Ideal Transformer
Ideal Transformer: Power Input = Power Output
Power is measured in watts
1watt = 1ampre * 1volt
P = I * V
Where:
P = power
I = current
V = Volts
Remember:
Pin = Pout
Therefore:
Vin * Iin = Vout * Iout
In Reality:
Pin = efficiency constant * Pout
Some number between zero and one
Loss due to heat?
Rectifiers
https://www.electronics-tutorials.ws/diode/diode_6.html
Full Wave Rectifier Circuit
Full-wave Rectifier with Smoothing Capacitor
5uF Smoothing Capacitor
50uF Smoothing Capacitor
Note Capacitor values impact on Smoothing
V=Q*C
Volts = charge (coulombs) * capacitance (Farads)
High Frequency Production and Rectification
5uF Smoothing Capacitor
50uF Smoothing Capacitor
Step 5
Step 1
Step 6
Step 2
Step 7
Step 4
Step 3
Step 8
Low Frequency
Rectification
Hi Frequency
Rectification
Production of
Hi Frequency
Voltage Ripple
Switches
High-Frequency Inverter on primary side of HV transformer because of low Voltage
Alternately Grid Control can stop x-ray production through biasing the grid (-4,000 V) to stop electron acceleration to target anode.
Grid Control is the fastest switching method
Timer(s)
Old equipment used mechanical timers
New equipment uses:
digital timers
Automatic Brightness Control (ABC, used in spot film devices)
Automatic Exposure Control (AEC, Fluoroscopy devices) also called Photo timer
Back-up Count Down Timer
Used in the case that the ABC or digital timer fail so exposure is terminated
Fluoroscopy Systems have another timer to warn user that the tube on time has increased (more on this in Fluoroscopy section)
ABC and AEC automatically adjust kV and mAs to achieve correct exposure to detector for optimal image
Phototimer/Automatic Exposure Control (AEC)
AEC detectors measure the radiation incident on the detector and terminate the exposure according to a preset optical density or signal-to-noise ratio achieved in the analog or digital image.
Chest cassette stand and the locations of ionization chamber detectors are shown. The desired signal to the image receptor and thus the signal-to-noise ratio may be selected at the operator’s console with respect to a normalized reference voltage.
Anti-Scatter Grid
Power Ratings, Heat Loading & Cooling
Power = Volts times Amperes (P= VI)
The Power rating (two ways to achieve)
= maximum power an x-ray generator can deliver
or
= maximum power a focal spot can accept
Testing is done for 0.1 seconds @ 100 kV
Example: a generator that can deliver 800 mA of current @ 100kV in 0.1 seconds has a power rating of:
0.8 Amperes X 100000 volts = 80,000 Watts or 80kW
When recommending an x-ray generator type the intended use should be taken into consideration. Power ratings can help. Duty cycle is many times required before recommendation can be made.
Heat Loading
The Heat Unit
The Heat Unit is labeled HU this is an unfortunate label because the CT x-ray linear attenuation units are translated into Hounsfield Units also labeled HU
Energy is in Joules where 1 Joule = 1 Watt * 1 Second
HU (Single Phase) = Voltspeak * Amperes * Seconds
HU (Single Phase) = kVpeak * mA * Seconds
HU (Three Phase) = 1.35 * (kVpeak * mA * Seconds)
HU (High Frequency) = 1.4 * (kVpeak * mA * Seconds)
Energy (Joules) = VoltsRMS * Amperes * Seconds
Energy (HU) = 1.4 * Energy (Joules)
Danger Ambiguation Situation
Danger Notice Vpeak & VRMS are used here
Anode Heating and Cooling Chart
Energy (Watts), HU (Joules), Joules = Watt Seconds
Therefore: Energy = HU/s = Volt Amps = Watts
CT scanner anode heating and cooling curves. Power input curves (2 kW 28kW) determined by kV and mA settings during continuous x-ray tube operation over time. Cool down curve shows the rate of cooling and indicates faster cooling at higher anode heat loads (temperature). In this example the maximum capacity is 5.7 MJ. For low power inputs, heating and cooling rates eventually equilibrate and reach a steady state, as shown for the 2, 4, and 8 kW curves.
4kW
Quesiton:
5.7 MJ = how many MHU?
Factors Effecting X-Ray Emissions
Anode Target Material
Tube Voltage
Tube Current
Beam Filtration
Generator Waveform
Anode Target Material
Higher Z = more bremsstrahlung radiation
Characteristic x-rays depend on target material
Tube Voltage
Tube Current
Beam Filtration: 21 CFR 1020.30
Designed Operating Range Measured x-ray tube voltage Minimum HVL (mm of Aluminum)
<51kV 30 0.3
40 0.4
50 0.5
51-70 kV 51 1.3
60 1.5
70 1.8
>70 kV 71 2.5
80 2.9
90 3.2
100 3.6
110 3.9
120 4.3
130 4.7
140 5.0
150 5.4
Spread Sheet Lookup table
Tube Filtration and Skin Entrance Dose
10cm PMMA (60kV) 20cm PMMA (80k) 30cm PMMA (100kV)
TUBE Current DOSE (mGy) TUBE Current DOSE (mGy) TUBE Current DOSE (mGy)
mAs %D Dose %D mAs %D Dose %D mAs %D Dose %D
3.8 0 264 0 6.8 0 1,153 0 14.5 0 4,827 0
5.0 32 188 -29 8.2 21 839 -27 16.5 14 3,613 -25
6.2 63 150 -44 9.3 37 680 -41 17.6 21 2,960 -39
8.8 132 123 -53 11.2 65 557 -52 19.8 37 2,459 -49
Filtration
0 mm Al
2 mm Al
0.1 mm Cu + 1 mm Al
0.2 mm Cu + 1 mm Al
OUTPUT DIGITAL IMAGE EQUIVALENT SIGNAL . TUBE FILTRATION, MEASURED CHANGES IN mAs, AND SURFACE DOSE (?Gy)
Filtration Causes:
1) acquisition technique much higher,
2) entrance dose much lower
3) no loss of image quality for digital imaging
Used Photo Timed Mode
Generator Wave Form
Affects the quality of the emitted x-ray spectrum. For the same kV, a single-phase generator provides a lower average potential difference than does a three-phase or high-frequency generator. Both the quality and quantity of the x-ray spectrum are affected
Output intensity (bremsstrahlung) spectra for the same tube voltage (kV) and the same tube current and exposure time (mAs) demonstrate the higher effective energy and greater output of a three-phase or high-frequency generator voltage waveform (?5% voltage ripple), compared with a single-phase generator waveform (100% voltage ripple)
Read more
Applied Sciences
Architecture and Design
Biology
Business & Finance
Chemistry
Computer Science
Geography
Geology
Education
Engineering
English
Environmental science
Spanish
Government
History
Human Resource Management
Information Systems
Law
Literature
Mathematics
Nursing
Physics
Political Science
Psychology
Reading
Science
Social Science
Home
Homework Answers
Blog
Archive
Tags
Reviews
Contact
google+twitterfacebook
Copyright © 2021 HomeworkMarket.com
Recent Comments